Toggle Main Menu Toggle Search

Open Access padlockePrints

A Comparison of 2D and 3D Edge Detectors in Semi Automated Measurements of Chamber Volumes Using 3D Echocardiographic Laboratory Phantom Images

Lookup NU author(s): Kun Wang, Dr Andrew SimsORCiD, Emeritus Professor Alan MurrayORCiD

Downloads


Abstract

Chamber volume measurement is fundamental for assessing many indices of cardiac function. 3D echocardiography offers great potential for accurate measurement, but many automated or semi-automated techniques suffer from acoustic noise and artefact, which can be reduced by suitable choice of edge detectors. We compared the effects of 2D and 3D first order derivative edge detectors with similar size on accuracy and repeatability in six volume measurement methods of 3D echo images of a laboratory balloon phantom. Overall accuracy compared with known volumes was better for 3D (-1.0ml; -4.0ml to 2.0ml 95% CI; P=0.53) than 2D (-11.6ml; -14.8ml to -8.5ml 95% CI; P<0.001). Repeatability was better for 3D (SD=0.9ml) than 2D (SD=1.5ml) with difference in coefficient of variation of 0.6 (P<0.001). 3D operators show potential for better performance with 3D echo images.


Publication metadata

Author(s): Wang K, Sims AJ, Murray A

Publication type: Conference Proceedings (inc. Abstract)

Publication status: Published

Conference Name: 36th Annual Computers in Cardiology Conference (CINC)

Year of Conference: 2009

Pages: 1-4

Date deposited: 07/09/2010

ISSN: 0276-6547

Publisher: IEEE

URL: http://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=5445484

Library holdings: Search Newcastle University Library for this item

ISBN: 9781424472819


Share