Toggle Main Menu Toggle Search

Open Access padlockePrints

The clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: a cohort study

Lookup NU author(s): Professor Tom Joyce, Professor Peter Blain, Dr Robert Jefferson

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Abstract Objective To determine whether elevated blood cobalt (Co) concentrations are associated with early failure of metal-on-metal (MoM) hip resurfacings secondary to adverse reaction to metal debris (ARMD). Design Cohort study. Setting Single centre orthopaedic unit. Participants Following the identification of complications potentially related to metal wear debris, a blood metal ion screening programme was instigated at our unit in 2007 for all patients with Articular Surface Replacement (ASR) and Birmingham MoM hip resurfacings. Patients were followed annually unless symptoms presented earlier. Symptomatic patients were investigated with ultrasound scan and joint aspiration. The clinical course of all 278 patients with ‘no pain’ or ‘slight/occasional’ pain and a Harris Hip Score greater than or equal to 95 at the time of venesection were documented. A retrospective analysis was subsequently conducted using mixed effect modelling to investigate the temporal pattern of blood Co levels in the patients and survival analysis to investigate the potential role of case demographics and blood Co levels as risk factors for subsequent failure secondary to ARMD. Results Blood Co concentration was a positive and significant risk factor (z=8.44, p=2×10–16) for joint failure, as was the device, where the Birmingham Hip Resurfacing posed a significantly reduced risk for revision by 89% (z=−3.445, p=0.00005 (95% CI on risk 62 to 97)). Analysis using Cox-proportional hazards models indicated that men had a 66% lower risk of joint failure than women (z=−2.29419, p=0.0218, (95% CI on risk reduction 23 to 89)). Conclusions The results suggest that elevated blood metal ion concentrations are associated with early failure of MoM devices secondary to adverse reactions to metal debris. Co concentrations greater than 20 µg/l are frequently associated with metal staining of tissues and the development of osteolysis. Development of soft tissue damage appears to be more complex with females and patients with ASR devices seemingly more at risk when exposed to equivalent doses of metal debris.


Publication metadata

Author(s): Langton DJ, Sidaginamale R, Joyce TJ, Natu S, Blain P, Jefferson RD, Rushton S, Nargol AVF

Publication type: Article

Publication status: Published

Journal: British Medical Journal Open

Year: 2013

Volume: 3

Issue: 3

Pages: 1-13

Print publication date: 12/03/2013

ISSN (electronic): 2044-6055

Publisher: BMJ Publishing Group

URL: http://dx.doi.org/10.1136/bmjopen-2012-001541

DOI: 10.1136/bmjopen-2012-001541


Altmetrics

Altmetrics provided by Altmetric