A device for minimizing the risk of overexposure of patients undergoing phototherapy

Lookup NU author(s): Emeritus Professor Brian Diffey

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Phototherapy irradiation cabinets normally rely on one timer or dose meter in order to control the ultraviolet (UV) exposure of patients. Faults may occur in the electronic control, or the operator may set the prescribed dose incorrectly. If this happens, a patient will receive an exposure different from that intended, which in the case of overexposure may result in severe burning resulting in hospitalization. Some phototherapy equipment has secondary devices (e.g. mechanical timer) for controlling the dose in case the primary system fails. We have designed and built a secondary dose meter that sounds an alarm if an overdose is about to occur. The unique feature of our device is that it measures the irradiance level within the cabinet and uses this value in combination with the dose set by the operator, to calculate the alarm time. Using the measured irradiance value in this manner ensures that the calculated exposure time compensates for variations in UV output as the lamps age.


Publication metadata

Author(s): Allan W, Diffey BL

Publication type: Article

Publication status: Published

Journal: Photodermatology, Photoimmunology & Photomedicine

Year: 2002

Volume: 18

Issue: 4

Pages: 199-201

ISSN (print): 0905-4383

ISSN (electronic): 1600-0781

Publisher: Wiley-Blackwell Publishing Ltd.

URL: http://dx.doi.org/10.1034/j.1600-0781.2002.00764.x

DOI: 10.1034/j.1600-0781.2002.00764.x


Altmetrics

Altmetrics provided by Altmetric


Share